SIRIO S2/T WITH SPIROMETER
The SIRIO S2/T is a medical device intended for the ventilation of adult, paediatric and neonatal patients (minimum 3 kg), designed for use in healthcare facilities, for emergency medical care, and during transport both within and outside hospitals. The adjustable respiratory parameters include pressure limit, respiratory rate, PEEP, minute volume and FiO2, allowing ventilatory support to be tailored to the patient’s needs.
MAX. FLOW
WEIGHT
EASY TO APPLY
ADDITIONAL FEATURES
STANDARDS COMPLIANT
- EN 60601-1:2006/A1:2013
- EN 60601-1-2:2015
- IEC 60601-1-6:2013
- EN 60601-1-8:2007/A11:2017
- EN 62304:2006/AC:2008
- EN 794-3:1998+A2:2009
- DIR. 2011/65/CE
- D.Lgs 49/2014
- DIR. 2011/65/UE RoHS 2 and subsequent modifications and integrations
- DIR. Delegated UE 2015/863 RoHS 3 and subsequent modifications and integrations
SUPPLY FEATURES
|
Electric power supply |
12Vdc or feeder: 100 – 240 Vac / 12Vdc |
|
Power consumption |
9 W |
|
Current consumption |
0,04 A – 220 Vac / 0,8 A – 12Vdc |
|
Standard internal battery |
1 Pb battery |
|
Internal battery operation (standard)(1) |
6,30 hours (Tolerance ±15%) |
|
Battery re-charging time (standard) |
8 hours |
(1) Battery life is indicative and measured in AUT-AST mode with the following settings: pressure 10 cmH2O, PEEP OFF, rate 15 bpm, tidal volume 9 l/min.
LUNG VENTILATOR FUNCTIONAL FEATURES
|
Type of ventilation |
IPPV volumetric – IPPV time cycled with PRESSURE SUPPORT |
|
Control modality |
Electronic |
|
Pressure automatic compensation (altitude) |
Present (max. 5000 mt) |
|
Respiratory parameters default setting |
The trigger is reset every time the device is turned off. Default – 1cmH2O |
|
Ventilation modes |
AUT + AST / PSV + APNOEA BACK UP / PEEP / CPAP |
|
Breathing rate |
From 5 to 70 bpm |
|
Max-Min Inspiratory Time / Expiratory Time (4) |
Insp Time: from 0,343 to 4,8 Exp Time: from 0,514 to 7,2 |
|
Tidal volume |
From 15 to 3000 ml |
|
Minute volume |
From 1 to 16 l/min |
|
I:E ratio |
1:1.5 |
|
Inspiratory pause |
Automatic setting during PSV mode |
|
Pressure support ventilation |
Adjustable from 0 (SPONT.) to 50 cmH2O |
|
Pressure limit |
Adjustable from 0 to 50 cmH2O |
|
PEEP |
Adjustable from 0 to 20 cmH2O |
|
Flow trend |
Constant and decreasing (PSV) |
|
O2 concentration |
50% O2 or 100% O2 |
|
Trigger detective method |
Through pressure transducer |
|
Pressure trigger ( I ) |
Adjustable from -1 to -5 cmH2O |
|
Inspiratory flow (FLOW) |
Max 40 l/min |
|
Patient circuits |
Single hose with exhalation valve |
Note: (4) Determined by Respiratory Rate.
PARAMETERS
|
Adjustable physiological respiratory parameters |
Pressure limit, respiratory rate, PEEP, Minute volume and FiO2 |
|
Measured parameters |
Respiratory rate (range: 0 ÷ 200 bpm) Rate accuracy ± 1 Bpm PAW: peak, mean, plateau, PEEP (range -20 ÷ 80 cmH2O) Pressure accuracy ± (2 cmH2O + 4% of the reading). Tidal Volume: Vte (range: 15 ÷ 3000 ml) Minute Volume (range: 1 ÷16 l/min) Volume Accuracy ± 20% for volumes> 50ml Expired Volume Accuracy ± (20% or ± 15ml) whichever is greater for volumes ≤50ml |
|
Displayed parameters |
Breathing cycle corrected by lungs icon, Breathing Rate (RATE), Expired Tidal Volume (VT), Minute Volume (VM), Instant Airways Pressure and Peak Airways Pressure (PAW), Mean Pressure (MAP), End Expiration Pressure (PEEP) |
Environmental Conditions
|
Operation |
|
|
Relative humidity |
15 – 95% non-condensing |
|
Temperature |
from -10 to +40°C |
|
Atmospheric pressure |
600hPa -1200hPa |
|
Storage |
|
|
Relative humidity |
< 95% non-condensing |
|
Temperature |
from -25 to +70°C |
|
Atmospheric pressure |
200hPa -1200hPa |
ALARMS
|
Remaining battery level from 100% to 51% |
Green LED |
|
Remaining battery level from 50% to 26% |
Yellow LED |
|
Remaining battery level from 25% to 0% |
Red LED |
|
O2 Supply |
Low gas supply pressure |
|
High PAW |
High airway pressure. Higher than 55 cmH2O |
|
Low PAW / Apnea |
Low airway pressure. Lower than 8 cmH2O |
|
Power failure |
Indicates that the device is disconnected from the mains supply |
FASTENING SYSTEMS REQUESTED
Quick-release 10G plate for wall mounting of the tubular support.
- Dimensions: 21 x h40 x 10 cm.
- Weight: 1,7 kg.
- Material: Lightweight aluminium alloy.
Tubular support for transporting a mechanical ventilator and a 2 L oxygen cylinder, with a Cordura bag.
- Dimensions: 45 x h41 x 35 cm.
- Weight: 3,9 kg.
- Material: Steel.